Pennsylvania Code & Bulletin
COMMONWEALTH OF PENNSYLVANIA

• No statutes or acts will be found at this website.

The Pennsylvania Code website reflects the Pennsylvania Code changes effective through 53 Pa.B. 8238 (December 30, 2023).

25 Pa. Code § 228.69. Termination of irradiation by the dose monitoring system or systems.

§ 228.69. Termination of irradiation by the dose monitoring system or systems.

 (a)  A dose monitoring system shall be capable of independently terminating irradiation.

 (b)  A primary system shall terminate irradiation when the preselected number of dose monitor units has been detected by the system.

 (c)  A secondary dose monitoring system shall terminate irradiation when either 110% of the preselected number of dose monitor units or 10 dose monitor units (whichever is greater) has been detected by the secondary dose monitoring system.

 (d)  For new equipment, an indicator on the control panel shall show which dose monitoring system has terminated irradiation.

Source

   The provisions of this §  228.69 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.



No part of the information on this site may be reproduced for profit or sold for profit.


This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.